6-2 Maintenance
RC04-PCR-LA
6.2
Before Requesting a Repair
If a problem occurs in the RC04-PCR-LA, check that problem in
accordance with the following Table 6-1. Also, disconnect the
Remote Controller from the PCR-LA AC power supply and check if
the power supply alone operates normally. If it is not possible to
recover a normal status, contact Kikusui distributor/agent to request
repairs.
Table 6-1 Troubleshooting Check Sheet
Problem
Check item
Results
Possible causes
LCD display
does not appear,
or display content
is not normal.
Check if the remote con-
troller cable has been
correctly connected.
NO
Improper remote con-
troller cable connection
(see 2.4)
Check the contrast
adjustment.
NO
Improper contrast
adjustment (see 3.1)
The Remote Con-
troller cannot be
operated.
Check if the remote con-
troller cable has been
correctly connected.
NO
Improper remote con-
troller cable connection
(see 2.4)
Check if the PCR-LA
AC power supply dis-
play unit indicates
“GPIB” or “RS-232C.”
YES
The PCR-LA AC power
supply is under control
of the GPIB Interface or
RS-232C Interface.