SYSTem Command
KIKUSUI Electronics Corp.
PWR-01 Interface Manual
SYSTem Command
SYST:BEEP:STAT
Turns the buzzer on and off.
Command
SYSTem:BEEPer:STATe <boolean>
SYSTem:BEEPer:STATe?
Parameter
Value:
ON(1)
Buzzer on (default)
OFF(0) Buzzer off
(Example)
SYST:BEEP:STAT OFF
Response
Returns the buzzer setting in NR1 format in response to SYST:BEEP:STAT?.
SYST:COMM:RLST
Sets the operation of the PWR-01 to local or remote.
Command
SYSTem:COMMunicate:RLSTate <character>
SYSTem:COMMunicate:RLSTate?
Parameter
Value:
LOCal
Sets the PWR-01 to local mode (Remote Disable; the RE-
MOTE LED turns off).
Disable state enables both panel operations and commands.
This is a substitute command for IEEE488.1 ren FALSE (Re-
mote Disable).
If the communication monitoring timer is set, set the timer to
off (OUTP:PROT:WDOG 0) before switching to local mode.
REMote Sets the PWR-01 operation to remote mode.
All panel keys except the LOCAL key are locked.
This is a substitute command for the IEEE488.1 REN (Remote
Enable) command and address designation.
RWLock Sets the PWR-01 operation to remote mode.
All panel keys are locked (the LOCAL key is also locked). This
is a substitute command for the IEEE488.1 llo (Local Lock
Out) command.
(Example)
SYST:COMM:RLST REM
Response
Returns the operation status of the PWR-01 in character format in response to the
SYST:COMM:RLST?.