Tutorial
KIKUSUI Electronics Corp.
PWR-01 Interface Manual
User code
You can execute sequences that you configured from the panel. You cannot view
the content of sequences from the panel, but you can view the user code.
To prevent unintentional sequence execution, we recommend that you set unique
user codes for each sequence and check the user code before executing the
sequence.
■
Setting step conditions
The parameters you set in a step are as follows: Use <n> of commands to specify
the step number.
You cannot set the output to on or off in a step. If you want to turn off the output, set
the voltage to 0 V and the current to 0 A.
When all the processes of a program are complete, the output is in the last-step
state.
Parameter
Description
Command
Default
Step time
Time to execute the step
PROG:STEP<n>:DWEL
1 s
Voltage
—
PROG:STEP<n>:VOLT
0 V
Voltage transition Voltage step transition
IMMediate
Current
—
PROG:STEP<n>:CURR
MAXimum
Current transition Current step transition
IMMediate
Trigger output
Trigger output on/off
PROG:STEP<n>:TRIGOUT Off
Trigger input
Pause until a trigger signal is
received
PROG:STEP<n>:TRIGIN
Off
Transition
Set the step transition. If you select IMMediate, the step changes in a step. If you
select RAMP, the step changes with a slope (actually, step transition at 100 ms
intervals) over the step time.
Example: Current value: 10 A
Immediate transition (IMMediate)
Step time: 1 s
Example: Current value: 10 A
Ramp transition (RAMP)
Step time: 1 s
1 s
10 A
1 s
10 A
Trigger output
Set the trigger output to on or off. When set to on, a trigger signal (100 µs) is
output from pin 7 of the TRIG OUT terminal on the rear panel simultaneously with
execution of the step.
Step n-1
Step n
TRIG OUT
Example: Setting the trigger output
at step n (TRIG ON)
100 µs
Trigger input
Set the step pause. When set to on, the sequence operation pauses before
executing the step. To release the pause state, enter a trigger signal (pin 7 of the
TRIG IN terminal on the rear panel).
Trigger signal input
(TRIG IN)
Paused state
Program
Resume execution
Trigger input turned on
at step n
Example: Pausing a sequence by turning on the trigger input and
releasing by applying a trigger input signal (TRIG IN)