[SOURce:]CURRent Command
KIKUSUI Electronics Corp.
PWR-01 Interface Manual
CURR:EXT:RANG
Sets the CC and CV control range that is used during external control.
This command is invalid while a sequence is running.
CURR:EXT:RANG and VOLT:EXT:RANG are aliases.
Command
[SOURce:]CURRent:EXTernal:RANGe <character>
[SOURce:]CURRent:EXTernal:RANGe?
Parameter
Value:
LOW
A range of 0 V to 5 V is used (default)
HIGH
A range of 0 V to 10 V is used.
When you send the VOLT:EXT:RANG command, this setting is set to the VOLT:EX-
T:RANG value.
(Example)
CURR:EXT:RANG HIGH
Response
Returns the CC and CV control range in character format in response to CURR:EX-
T:RANG?.
CURR:EXT:SOUR
Sets whether constant current will be controlled externally.
This command is invalid while a sequence is running.
Command
[SOURce:]CURRent:EXTernal:SOURce <character>
[SOURce:]CURRent:EXTernal:SOURce?
Parameter
Value:
NONE
External control is not performed (default).
VOLTage External control is performed.
Settings are reset to default when the *RST command is sent.
(Example)
CURR:EXT:SOUR VOLT
Response
Returns whether constant current is being controlled externally in character format
in response to CURR:EXT:SOUR?.