[SOURce:]VOLTage Command
KIKUSUI Electronics Corp.
PWR-01 Interface Manual
VOLT:SST:RISE
Sets the soft start time when the output is turned on.
This command is invalid when the startup state when the output is turned on is set
to CC priority (SYST:CONF:STAR:PRI CC).
Command
[SOURce:]VOLTage:SSTart:RISE <numeric>
[SOURce:]VOLTage:SSTart:RISE?
Parameter
Value:
0
Disable soft start (default)
0.5 to 10.0
Soft start time
A SCPI error (-222, “Data out of range”) occurs if outside the range.
Unit:
S
Settings are reset to default when the *RST command is sent.
(Example)
VOLT:SST:RISE 2500MS
Response
Returns the soft start time in NR3 format in response to VOLT:SST:RISE?.
VOLT:TRIG
Sets the voltage that is applied when a trigger is sent.
When the PWR-01 is set so that constant voltage is controlled externally (VOLT:EX-
T:SOUR VOLT), settings performed by this command are invalid.
This command is invalid while a sequence is running.
Command
[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude] <numeric>
[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude]?
Parameter
Value:
command
Unit:
V
Settings are reset to default when the *RST command is sent.
When you send the VOLT command, this setting is set to the VOLT value.
(Example)
VOLT:TRIG 120
Response
Returns the voltage value that is applied when a trigger is received in NR3 format
in response to VOLT:TRIG?.