OUTPut Command
KIKUSUI Electronics Corp.
PWR-01 Interface Manual
OUTP:PROT:CLE
Clears alarms.
Command
OUTPut:PROTection:CLEar
OUTP:PROT:WDOG
Sets the Communication monitor timer.
When an alarm occurs, set the value to zero first and then clear the alarm.
Command
OUTPut:PROTection:WDOG[:DELay] <numeric>
OUTPut:PROTection:WDOG[:DELay]?
Parameter
Value:
0
Communication monitoring function off (default)
1, 3, 10, 30, 100, 300, 1000, 3000
A SCPI error (-222, “Data out of range”) occurs if outside the range.
If a value that cannot be set is specified, the next higher value is set.
(Example) If OUTP:PROT:WDOG 31 is specified, 100 is set.
Unit:
S
(Example)
OUTP:PROT:WDOG 30
Response
Returns the Communication monitor timer in NR1 format in response to OUT-
P:PROT:WDOG?.
Related Command
OUTP:PROT:CLE