IEEE488.2 Common Commands
KIKUSUI Electronics Corp.
PWR-01 Interface Manual
*ESE
Sets the event status enable register that is counted by the event summary bit (ESB)
of the status byte.
Command
*ESE <NRf>
*ESE?
Parameter
Value: 0 to 255
(Example) When *ESE 16 is transmitted, bit 4 of the event status enable register is
set. Each time the execution error bit (bit 4) of the event status register is set, the
summary bit (ESB) of the status byte is set.
Response
Returns the value of the event status enable register in NR1 format in response to
*ESE?.
*ESR
Queries the event status register.
Registers that are read are cleared.
Command
*ESR?
Response
Returns the value of the event status register in NR1 format in response to *ESR?
and clears the register.