[SOURce:]VOLTage Command
KIKUSUI Electronics Corp.
PWR-01 Interface Manual
VOLT:PROT
Sets the overvoltage protection (OVP).
This command is invalid while a sequence is running.
Command
[SOURce:]VOLTage:PROTection[:LEVel] <numeric>
[SOURce:]VOLTage:PROTection[:LEVel]?
Parameter
Value:
10 % to 112 % of the rated output voltage (OVP; MAXimum by default)
A SCPI error (-222, “Data out of range”) occurs if outside the range.
Unit:
V
Settings are reset to default when the *RST command is sent.
If you enable the voltage limit (VOLT:LIM:AUTO ON) when the voltage setting is
higher than 95 % of the OVP trip point, the OVP trip point setting is changed to 105
% of the voltage setting.
(Example)
VOLT:PROT 48
Response
Returns the OVP value in NR3 format in response to VOLT:PROT?.
Related Command
OUTP:PROT:CLE
VOLT:LIM:AUTO
VOLT:SST:FALL
Sets the soft stop time when the output is turned off.
This command is invalid when the startup state when the output is turned on is set
to CC priority (SYST:CONF:STAR:PRI CC).
Command
[SOURce:]VOLTage:SSTart:FALL <numeric>
[SOURce:]VOLTage:SSTart:FALL?
Parameter
Value:
0
Disable soft stop (default)
0.5 to 10.0
Soft stop time
A SCPI error (-222, “Data out of range”) occurs if outside the range.
Unit:
S
Settings are reset to default when the *RST command is sent.
(Example)
VOLT:SST:FALL 2500MS
Response
Returns the soft stop time in NR3 format in response to VOLT:SST:FALL?.