[SOURce:]CURRent Command
KIKUSUI Electronics Corp.
PWR-01 Interface Manual
CURR:PROT:DEL
Sets the detection time of OCP activation.
This command is invalid while a sequence is running.
Command
[SOURce:]CURRent:PROTection:DELay <numeric>
[SOURce:]CURRent:PROTection:DELay?
Parameter
Value:
0.0 to 2.0 Detection time of OCP activation. (0 s by default)
A SCPI error (-222, “Data out of range”) occurs if outside the range.
Unit:
S
(Example)
CURR:PROT:DEL 1.5
Response
Returns the detection time of OCP activation in NR3 format in response to CUR-
R:PROT:DEL?.
Related Command
CURR:PROT
CURR:SST:FALL
Sets the soft stop time when the output is turned off.
This command is valid when the startup state when the output is turned on is set to
CC priority (SYST:CONF:STAR:PRI CC).
Command
[SOURce:]CURRent:SSTart:FALL <numeric>
[SOURce:]CURRent:SSTart:FALL?
Parameter
Value:
0
Disable soft stop (default)
0.5 to 10.0
Soft stop time
A SCPI error (-222, “Data out of range”) occurs if outside the range.
Unit:
S
Settings are reset to default when the *RST command is sent.
(Example)
CURR:SST:FALL 2500MS
Response
Returns the soft stop time in NR3 format in response to CURR:SST:FALL?.