OUTPut Command
KIKUSUI Electronics Corp.
PWR-01 Interface Manual
OUTP:EXT
This is an old style command.
When creating new programs, use OUTP:EXT:ADV.
Command
OUTPut:EXTernal[:STATe] <boolean>
OUTPut:EXTernal[:STATe]?
Settings are reset to default when the *RST command is sent.
Response
NR1 format
OUTP:EXT:ADV
Sets whether output will be turned on and off externally.
Set the external control logic with
This command is invalid while a sequence is running.
Command
OUTPut:EXTernal[:STATe]:ADVance <character>
OUTPut:EXTernal[:STATe]:ADVance?
Parameter
Value:
ON
External control is performed.
Disables the output-on/ off delay function and soft start
and soft stop functions.
ENHancedon External control is performed.
Enables the output-on/ off delay function and soft start
and soft stop functions.
OFF
External control is not performed (default).
Settings are reset to default when the *RST command is sent.
(Example)
OUTP:EXT:ADV ON
Response
Returns whether output will be turned on and off externally in CHAR format in re-
sponse to OUTP:EXT:ADV?.