2-8 Operating Procedure
PFX2000
2.3
Using the Trip Connector
You can externally shut off the power switch on the unit fame by using a trip con-
nector. The power switch is turned off when the positive and negative terminals of
the trip connector are shorted.
■
Wires and tools required for connection
1. Wires
Stranded wire: AWG22 to 20 (0.33 mm
2
to 0.52 mm
2
)
2. Flat-blade screwdriver
Use the screw driver that comes with the package.
3. Wire stripper
Wire stripper suitable for the wires described above.
■
Electrical speci
fi
cations
The electrical speci
fi
cations of the TRIP input are shown in Table 2-3. Use a
fl
oat-
ing contact signal or an open-collector output that meets these speci
fi
cations.
Table 2-3
Electrical Specifications of TRIP Input
•
Connection to the power connector must be performed while the power
switch is turned off.
•
Be sure the exposed section of the wire does not touch the chassis or the
wires of the adjacent connector.
• After connecting the wire, pull on it lightly to check that it does not come
loose.
Fig. 2-7
Trip Connector Connection
PFX2331
PFX2332
When opened
24 VDC
When shorted
150 mA
300 mA
Time of appli-
cation
10 s maximum
2 s maximum
CAUTION
Insert the wire with 10 mm of
the covering removed here.
Insert the wire while pressing
this section using a flat-blade
screwdriver.
10 mm