A-6 Appendix
PFX2000
■
Pulse current measurement
PFX2011
In pulse current measurement, each pulse current is measured and the average cur-
rent is calculated over a unit time.
IMn: Measured value of each pulse current
Tn: Time width of each pulse
Tave: Unit time for averaging (
fi
xed to 500 ms)
Below is an explanation using an actual waveform.
Fig. A-9 is an example for a pulse waveform with a pulse time width that is shorter
than the unit time for averaging.
In this case, the average current is derived by summing the product of the measured
pulse current and the pulse time width until the unit time is reached (numerator of
Equation 1). When the unit time is reached, the sum is divided by the unit time to
derive the average current.
Fig. A-9
Averaging Example 1 of Pulse Current Measurement
Fig. A-10 is an example for a pulse waveform with a pulse time width that is longer
than or approximately the equal as the unit time for averaging. Similar to Fig. A-9,
the measured value of the current that
fl
ows during the unit time for averaging is
used to calculate the average current.
Fig. A-10
Averaging Example 2 of Pulse Current Measurement
Average current =
IM1 • T1 + IM2 • T2 + IM3 • T3 + IM4 • T4
Tave
(Equation 1)
Pulse current
Current large
↑
Current small
↓
T1
IM1
IM2
T1
IM1
IM2
T1
IM1
IM2
T2
IM1
IM2
Average calculation
Unit time for averaging (500 ms)
T2
T1
Average current = IM2 • T2+IM1 • T1+IM2 • T2+IM1 • T1+ ••• +IM1 • T1/500 ms
I1
I2
I3
I4
I1
I2
Tave 1
Tave 2
Tave 3
Tave 4
Tave 5
Tave 6
Current large
↑
Current small
↓
Pulse current
Average current
Tave n: Unit time for averaging (fixed to 500 ms)