PFX2000
Operating Procedure 2-7
Fig. 2-5
Output Cable Connection
Connection when using the PFX2021
When using the PFX2021 Charge/Discharge Power Supply Unit, connect the output
cables to the odd-numbered output connectors on the unit frame. Even-numbered
output connectors cannot be used.
Fig. 2-6
Connection When Using the PFX2021
FRAM
E
DUT
CONT
TERM
N
TP-BUS
TRIP
+
-
BATTERY
TESTER F
RAME UN
IT
No.
KIKUSU
I ELEC
TRONICS
CORP.
OFF
ON
MADE IN J
APAN
1
2
3
4
5
6
7
8
9
10
AC LINE
IMPEDANC
E
VA MAX
1.5k
FREQUEN
CY
50
/
60 Hz
LINE V
OLTAGE
RANGE
100-120V
200-240V
IF THIS C
OLUMN I
S BLANK
,
THE UNIT
IS WIRED
IN 200V
SETTING SUP
PLY
STANDA
RD
– +
– -
– +S
– -S
– + T
– - T
– FG
OUTPUT
TERMINA
L
No. 1-10
0
E D
C
B
A
9
8
7
6
5
4
3
2
1
+
–
Do not leave an extra length of cable.
Fix the connector in place by
fastening the top and bottom
screws.
Current
wire
Voltage
sensing wire
Thermistor
Temperature
sensing wire
DUT (battery)
Battery holder
If terminals are available for voltage measurement,
connect the voltage sensing wires to them.
Combination of the PFX2332 Unit Frame and 5 PFX2021 Charge/Discharge Power Supply Units
Connect the output cables to the odd-numbered output connectors.