PFX2000
Appendix A-3
Fig. A-3
Correction Converter
Fig. A-4 is an example of the auto
fi
ne operation when the current is set to 3.000
0
A.
Fig. A-4
Auto Fine Operation Example
Operation
•
The auto
fi
ne function is active only during constant current (CC) operation.
During CC-CV operation, auto
fi
ne operates only during the CC period. It does
not operate during pulse charge/discharge operation.
•
The range of correction is limited to
±
30 digits. For example, if the difference
between the preset value and the measured value is 40 digits, correction is per-
formed up to 30 digits. In this case, the difference between the preset value and the
measured value is 10 digits. This limitation makes identi
fi
cation of abnormality eas-
ier when the accuracy of the settings is off by an extreme amount for some reason.
•
The preset value and the measured value never are completely the same. For
performance reasons, the operation stops after correcting the values within
approximately 3 digits.
•
Correction is performed at 1-s cycle. It may take from several seconds to several
tens of seconds for the correction to complete.
•
For the procedure in turning ON/OFF the auto
fi
ne function, see section 4.3.1,
“Module” in the
BPChecker2000 User's Manual
.
16 bit D/A converter
8 bit D/A converter
CC reference signal
Correction converter
Preset constant current value: 3.0000 A
Line corresponding to
the measured value of
3.0000 A
16-bit D/A converter
setting
8-bit D/A converter setting
(Correction operation)
Current