1-18 Setup
PFX2000
Fig. 1-9 shows an example in which the maximum number of PFX2021 Charge/
Discharge Power Supply Units is connected to a single control unit.
The maximum number of unit frames that can be connected to a single control unit
is 12. The PFX2021 has one channel per unit. Therefore, the maximum number of
channels that can be connected to a single control unit is 60. If the PFX2021 is
installed, only odd node numbers are valid.
If a system is con
fi
gured so that all of the charge/discharge power supply units are
PFX2021s, the maximum number of channels that the control unit can control is 60.
Fig. 1-9
Frame Addresses and Node Numbers
(Example of 60-Channel Connection Using PFX2021s)
node node node node node
51 53 55 57 59
node node node node node
41 43 45 47 49
node node node node node
31 33 35 37 39
node node node node node
21 23 25 27 29
node node node node node
11 13 15 17 19
node node node node node
01 03 05 07 09
node node node node node
111 113 115 117 119
node node node node node
101 103 105 107 109
node node node node node
91 93 95 97 99
node node node node node
81 83 85 87 89
node node node node node
71 73 75 77 79
node node node node node
61 63 65 67 69
PFX2121
Control unit
PFX2332 Unit Frame
5 PFX2021 Charge/Discharge Units
PFX2332 Unit Frame
5 PFX2021 Charge/Discharge Units
TP-BUS1
T P -BUS 2
Frame
address
6
Frame
address
5
Frame
address
4
Frame
address
3
Frame
address
2
Frame
address
1
Frame
address
C
Frame
address
B
Frame
address
A
Frame
address
9
Frame
address
8
Frame
address
7