S
pecifications
Voltmeter
Number of display digits (full scale)
300.0 Vrms
Accuracy
*6
Within
±
1 % of FS
Protection function
*17
Peak overcurrent protection (POCP)
*18
Approx. 48 Apeak
Overcurrent protection (OCP)
*19
Approx. 11.5 Arms
Overvoltage protection (OVP)
*18
Approx. 470 Vpeak
Overpower protection (OPP)
Approx. 1 150 W
Overheat protection (OHP)
*20
−
Internal circuit protection(FUSE BRK)
Internal fuse blown
External alarm 1 (EXT1)
Another unit has detected an alarm during
parallel or tracking operation.
External alarm 2 (EXT2)
Alarm detected when the PC01-PCZ1000A
flat cable comes unplugged during parallel or
tracking operation (only on slave units that
become disconnected from the master).
Input power (AC)
Voltage range (nominal value)
*21
90 Vrms to 110 (100) Vrms
108 Vrms to 132 (120) Vrms
180 Vrms to 220 (200) Vrms
216 Vrms to 250 (240) Vrms
Frequency
50 Hz/60 Hz
Power consumption (apparent power)
MAX 220 VA
*1.
Input-voltage range in which the rated input current can flow.
*2.
For an input voltage of 100 Vrms or greater, the maximum current is derated to the rated input
power (1 000 W).
*3.
For an input voltage of 100 Vrms or less, the maximum power is limited by the rated input
current (10 Arms).
*4.
Minimum input voltage at which the input current starts to flow
*5.
The input-current waveform does not change with changes in the input-voltage waveform.
The RMS value of the input current is kept constant (response rate: approximately 1 s).
(Response rate: Time required to reach
±
10% of the steady value (value reached 5 seconds
or more after state change))
*6.
At temperature(23
±
5
°
C)
*7.
Changes in the input current when variations in the rated-voltage range are given at an input
voltage of 100 Vrms and an input current of 10 Arms, based on the nominal value of the input
line voltage.
*8.
Changes in the input current when the input voltage is changed from 10 Vrms to 280 Vrms at
an input current of 3.57 Arms (rating at an input voltage of 280 Vrms).
*9.
The input-current waveform does not change with changes in the input-voltage waveform.
This mode results in an input current (RMS value) that is proportional to the RMS value of the
input voltage (response rate: approximately 1 s).
*10. S represents the unit of conductance (siemens).
Conductance [S] = 1/resistance value [
Ω
]
Conductance [S] × input voltage [V] = load current [A]
*11. At an input voltage of 100 Vrms.
*12. Changes in the resistance value when the input voltage is varied from 10 Vrms to 100 Vrms
at an input current of 0.5 A or more.
*13. The input-current waveform does not change with changes in the input-voltage waveform.
This mode results in an input current (RMS value) that is inversely proportional to the RMS
value of the input voltage (response rate: approximately 1 s).
*14. At an input voltage of 100 Vrms
*15. Changes in the power value when the input voltage is varied from 10 Vrms to 100 Vrms.
*16. Varies the angular width of the current at the approximate input-voltage peak, based on a
sinusoidal current waveform.
*17. Turns off the LOAD key when the protection function is activated.
*18. Turns off the LOAD key within 20 ms.
*19. Turns off the LOAD key within 3 ms.
*20. Detects the internal-heat-sink surface temperature to turn off the LOAD key.
*21. Selectable by Switch position.
Содержание PCZ1000A
Страница 4: ...ii PCZ1000A This page is intentionally blank ...
Страница 14: ...xii PCZ1000A Front panel 18 17 Control panel Control panel 1 Display Display 2 3 4 5 6 7 8 9 10 11 12 13 14 ...
Страница 17: ...General Description This chapter gives an overview and describes the features of the PCZ1000A ...
Страница 35: ...Operation This chapter describes each mode and function of the operation ...
Страница 82: ...5 4 PCZ1000A This page is intentionally blank ...
Страница 83: ...Specifications This chapter gives the specifications and dimensions of the PCZ1000A ...
Страница 88: ...6 6 PCZ1000A This page is intentionally blank ...
Страница 89: ...The appendix contains the basic operating modes operating range and troubleshooting Appendix ...
Страница 99: ......