3-28
PCZ1000A
Connecting Units for Tracking Operation
Risk of electric shock
• Turn off the POWER switch when you are touching the load input
terminals, PARALLEL IN connector, or PARALLEL OUT connector.
Also, re-attach the load input terminal cover after connecting the
cables.
• Always fit the protective socket when not using the PARALLEL IN
connector or PARALLEL OUT connector.
Risk of damage to the PCZ1000A.
• Take care not to get the PARALLEL IN connector and PARALLEL OUT
connector mixed up.
• Ensure that the load cable is as short as possible with a sufficient cross
section for the current level being used.
• Ensure that the insulation on the load cable you use has good fire resistance.
• Ensure that the cables to the load input terminals on each of the PCZ1000A are
as similar in length as possible.
1
Turn off the POWER switch.
2
Connect the cables to the load input terminals.
Referring to the connection diagram in Fig.3-10, connect the cables to the load input
terminals on each unit (which must all be of the same model).
3
Connect the parallel operation connector.
Use the option flat cable (PC01-PCZ1000A) to connection between the PARALLEL
OUT and PARALLEL IN connectors on each unit.
Procedure for Setting Master and Slave Units
1
Turn on the POWER switch on each unit.
Either turn on the POWER switch on all units at the same time or turn on the master
unit first.
• If the power to the slaves is turned on first, turn on the POWER switch of the
master unit within 30 seconds. An external alarm 2 occurs if this time is longer
than 30 seconds.
2
Press the LOAD key to go to the load-off state.
3
Press SHIFT + I SET(CONFIG).
4
Use the ENTER key to select ">1:Norm /M/S".
WARNING
CAUTION
Содержание PCZ1000A
Страница 4: ...ii PCZ1000A This page is intentionally blank ...
Страница 14: ...xii PCZ1000A Front panel 18 17 Control panel Control panel 1 Display Display 2 3 4 5 6 7 8 9 10 11 12 13 14 ...
Страница 17: ...General Description This chapter gives an overview and describes the features of the PCZ1000A ...
Страница 35: ...Operation This chapter describes each mode and function of the operation ...
Страница 82: ...5 4 PCZ1000A This page is intentionally blank ...
Страница 83: ...Specifications This chapter gives the specifications and dimensions of the PCZ1000A ...
Страница 88: ...6 6 PCZ1000A This page is intentionally blank ...
Страница 89: ...The appendix contains the basic operating modes operating range and troubleshooting Appendix ...
Страница 99: ......