3-30
PCZ1000A
3.7.4
Calibration of the Ammeter on the Master Unit During
Parallel Operation
The ammeter of the PCZ1000A when used for parallel operation is calibrated at the
shipment from the factory. However, the ammeter accuracy can be improved further
by calibrating the instrument under the actual parallel operation conditions
Preparation for Calibration
•
To minimize the measurement error due to startup drift, allow the current to
flow for 30 minutes before starting the test. Also, maintain the ambient
temperature at 23
°
C ± 5
°
C.
•
Equipment
AC power supply: An AC power supply with adequate capacity to supply the
combined requirements of all the instruments.
Output current (RMS): Number of units x 10 A
Output current (peak): Number of units x 40 A
Output voltage (RMS): 10 V
AC ammeter:
A meter with sufficient range to measure an RMS current of number
of units x 10 A and a peak current of number of units x 40 A.
Accuracy 0.5 % or better, frequency: 45 Hz to 65 Hz.
Fig.3-11 Wiring Diagram for Calibration
PARALLEL IN
PARALLEL IN
PARALLEL OUT
PARALLEL OUT
L
L
L
L
N
N
N
G
G
G
N
G
AC ammeter
AC voltmeter
AC power
supply
Flat cable
Master
Slave
Slave
A
V
Содержание PCZ1000A
Страница 4: ...ii PCZ1000A This page is intentionally blank ...
Страница 14: ...xii PCZ1000A Front panel 18 17 Control panel Control panel 1 Display Display 2 3 4 5 6 7 8 9 10 11 12 13 14 ...
Страница 17: ...General Description This chapter gives an overview and describes the features of the PCZ1000A ...
Страница 35: ...Operation This chapter describes each mode and function of the operation ...
Страница 82: ...5 4 PCZ1000A This page is intentionally blank ...
Страница 83: ...Specifications This chapter gives the specifications and dimensions of the PCZ1000A ...
Страница 88: ...6 6 PCZ1000A This page is intentionally blank ...
Страница 89: ...The appendix contains the basic operating modes operating range and troubleshooting Appendix ...
Страница 99: ......