PCZ1000A
4-9
4
Remote Control
4.4
Assignment of Register Bits
A bit becomes "1" when the current state is "true".
The returned value is the sum (bitwise AND) of the bit decimal values.
Example:
The value returned when both the overheating protection function [4] and
overpower protection function [16] are activated is:
4 + 16 = 20
where 20 is response data.
Fault Register
The bits in this register are latched when the corresponding status is present and
cleared after the register is read to provide the response message to an "FAU?"
request. To determine the present status, send "FAU?" to clear the register and then
send "FAU?" again to retrieve the present status.
Fault Register 2
Bit
Decimal
Value
Status of Fault Register
0
1
Overvoltage protection triggered
(OVP)
1
2
Overcurrent protection or peak
overcurrent protection triggered
(OCP, POCP)
2
4
Overheat protection triggered
(OHP)
3
8
Internal circuit protection triggered (FUSE BRK)
4
16
Overpower protection triggered
(OPP)
5
32
CP-range operation
(CP)
6
64
CR-range operation
(CR)
7
128
CC-range operation
(CC)
Bit
Decimal
Value
Status of Fault Register 2
0
1
External alarm 1 triggered
(EXT1 ALARM)
1
2
-
-
2
4
-
-
3
8
-
-
4
16
-
-
5
32
-
-
6
64
-
-
7
128
-
-
Содержание PCZ1000A
Страница 4: ...ii PCZ1000A This page is intentionally blank ...
Страница 14: ...xii PCZ1000A Front panel 18 17 Control panel Control panel 1 Display Display 2 3 4 5 6 7 8 9 10 11 12 13 14 ...
Страница 17: ...General Description This chapter gives an overview and describes the features of the PCZ1000A ...
Страница 35: ...Operation This chapter describes each mode and function of the operation ...
Страница 82: ...5 4 PCZ1000A This page is intentionally blank ...
Страница 83: ...Specifications This chapter gives the specifications and dimensions of the PCZ1000A ...
Страница 88: ...6 6 PCZ1000A This page is intentionally blank ...
Страница 89: ...The appendix contains the basic operating modes operating range and troubleshooting Appendix ...
Страница 99: ......