lxx
Keysight E1412A Digital Multimeter User Guide
Keysight E1412A Multimeter Application Information
E1412A Multimeter Application Examples
CAL:ZERO:AUTO OFF
Turn off autozero (makes faster measurements).
TRIG:SOUR IMM
Set the trigger source to immediate.
TRIG:COUN 3
Set trigger count to 3.
SAMP:COUN 10
Set sample count to 10.
INIT
INITiate command puts multimeter in wait-for-trigger state;
internal trigger immediately occurs here and measurements are
stored in the multimeter's internal memory.
FETC?
Transfer measurements from the multimeter's internal memory to the
output buffer and retrieve them with the computer.
enter statement
Enter reading into computer.
Comments
–
The
CONFigure
command configures the multimeter for the function
specified. This
CONFigure
command specifies a range parameter of 18
(expected input is 18V; the multimeter sets a range to accommodate that
input which will be 100V). It does not initiate the measurement.
–
The autozero function is disabled to speed up the measurement process.
See the
CALibrate:ZERO:AUTO
command in the Command Reference for
more information.
–
Trigger source (
TRIG:SOUR
) is set for immediate internal triggers.
A trigger count (
TRIG:COUN
) of 3 is set; the multimeter will accept three
triggers.
–
The sample count (
SAMP:COUN
) is set for 10 samples per trigger.
–
The
INITiate
command puts the multimeter in the wait-for-trigger state.
The trigger source is “
IMMediate
” which specifies the internal trigger
source. This trigger occurs immediately and causes the measurement
process to begin. This will cause the multimeter to make 30
measurements; 10 samples for each of three internal triggers.
–
The
FETCh?
command causes the readings to be transferred to the output
buffer and you must provide the I/O construct to retrieve the readings and
enter them into the computer.