Keysight E1412A Digital Multimeter User Guide
li
Math Operations (CALCulate Subsystem)
Keysight E1412A Multimeter Application Information
–
The null value is adjustable and you can set it to any value between
0 and
120% of the highest range, for the present function.
–
Clearing the
NULL
value. The null value is stored in volatile memory; the
value is cleared when power is removed, after resetting the multimeter or
after a function change.
Two Ways to Store the NULL Offset Value
–
The null value is stored in the multimeter's Null Register. You can enter a
specific number into the null register using the
CALCulate:NULL:OFFSet <
value
>
command. Any previously stored value is
replaced with the new value. Use the following commands to activate the
NULL
function and input a null value. The calculate state must be enabled
before you can store a value in the Null Register.
CONF:<
function
>
Clears the null offset value.
CALCulate:FUNCtion NULL
Set math function to NULL.
CALCulate:STATe ON
Enable math operation.
CALCulate:NULL:OFFSet <
value
>
Store a null offset value.
–
Another way to enter the null value is to let the multimeter store the first
reading in the register. After you enable the
NULL
function with the
CALC:STATe ON
command, the first measurement you obtain will be zero
(if you have not stored a value as described in the previous bullet). The
measured value is stored as the
NULL
offset value and subtracted from
itself to result in the zero reading. All subsequent measurements will have
the offset value subtracted from them. If you previously stored a
NULL
offset value using
CALC:NULL:OFFS <
value
>
as in the commands in the
above bullet, the first reading does not overwrite the stored offset value
but returns with the previous offset value subtracted.
CONF:<
function
>
Clears the null offset value.
CALCulate:FUNCtion NULL
Set math function to NULL.
CALCulate:STATe ON
Enable math operation.
** Set up the system to generate the offset of concern (e.g., short
** input leads for 2-wire ohms measurements that will follow).
READ?
Measures and stores the offset value.
dB Measurements
Each dB measurement is the difference between the input signal and a stored
relative value, with both values converted to dBm.
dB = reading in dBm - relative value in dBm
–
Applies to dc voltage and ac voltage measurements only.
–
The relative value is adjustable and you can set it to any value between
0dBm and
200.00dBm (well beyond the multimeter's measurement
capabilities).
Содержание E1412A
Страница 1: ...Keysight E1412A 6 1 2 Digit Multimeter User Manual and SCPI Programming Guide 75000 Series C ...
Страница 2: ......
Страница 6: ...vi ...
Страница 18: ...xviii Keysight E1412A Multimeter User Guide ...
Страница 208: ...ccviii Keysight E1412A Digital Multimeter User Guide Multimeter Command Reference SCPI Command Quick Reference ...
Страница 242: ......