lii
Keysight E1412A Digital Multimeter User Guide
Keysight E1412A Multimeter Application Information
Math Operations (CALCulate Subsystem)
–
Clearing the relative value. The relative value is stored in volatile memory;
the value is cleared when power is removed, after the module is reset or
after a function change.
Storing the dB Reference Value
Do not confuse this operation with the dBm reference (DBM) function. See the
next section, “dBm Measurements”, and take note of the multimeter's reference
resistance setting (dB uses a reference level, dBm uses a reference resistance).
–
The dB reference value is stored in the multimeter's dB Relative Register.
You can enter a specific number into the register using the
CALCulate:DB:REFerence <
value
>
command. Any previously stored value
is replaced with the new value. Use the following commands to activate
the dBm function and input a reference value. The calculate state must be
enabled before you can store a value in the dB Relative Register.
CALCulate:FUNCtion DB
Set math function to DB.
CALCulate:STATe ON
Enable math operation.
CALCulate:DB:REFerence <
value
>
Store a dB reference value.
dBm Measurements
The dBm operation calculates the power delivered to a resistance referenced to 1
milliwatt.
dBm =
–
Applies to dc voltage and ac voltage measurements only.
–
You can choose from 17 different reference resistance values. The factory
setting for the reference resistance is 600
. Set your desired value with
the
CALC:DBM:REF <
value
>
command.
–
The choices for
<
value
>
are:
50
,
75
,
93
,
110
,
124
,
125
,
135
,
150
,
250
,
300
,
500
,
600
,
800
,
900
,
1000
,
1200
, or
8000
ohms.
–
The reference resistance is stored in nonvolatile memory, and does not
change when power is removed or after the multimeter is reset.
Storing the dBm Reference Resistance Value
Do not confuse this operation with the dB reference (
DB
) function. See the
previous section, “dB Measurements”, and take note of the multimeter's dB
reference setting (dB uses a reference level, dBm uses a reference resistance).
–
Use the following commands to activate the dBm function and input a
reference resistance value. The calculate state must be enabled before you
can store a value in the Reference Resistance Register.
10 log
10
reading
2
(reference resistance) (1 mW)
----------------------------------------------------------------------------