Keysight E1412A Digital Multimeter User Guide
liii
Math Operations (CALCulate Subsystem)
Keysight E1412A Multimeter Application Information
CALCulate:FUNCtion DBm
Set math function to DBm.
CALCulate:STATe ON
Enable math operation.
CALCulate:DBM:REFerence <
value
>
Store a dBm reference.
LIMit Function
The limit test operation enables you to perform pass/fail testing against limits
you specify using the
CALCulate:LIMit:UPPer
and
LOWer
commands.
–
Applies to all measurement functions.
–
You can set the upper and lower limits to any value between 0 and
120%
of the highest range, for the present function. The upper limit selected
should always be a more positive number than the lower limit. The default
upper and lower limits are both “0”.
–
The upper and lower limits are stored in volatile memory; the multimeter
sets both limits to 0 when power is removed from the multimeter, after the
multimeter is reset or after a function change.
–
You can configure the multimeter to generate a request for service (SRQ)
on the first occurrence of a failed reading. See the Status System Register
Diagram in Figure 3-1. Bits 11 and 12 of the Questionable Data Register
provide the high and low limit error signals that can be enabled in the
status byte to generate the request for service.
–
Use the following commands to activate the
LIMit
function and input upper
and lower limit values. The calculate state must be enabled before you can
store a value in the Upper Limit and Lower Limit Registers.
CALCulate:FUNCtion LIMit
CALCulate:STATe ON
CALCulate:LIMit:UPPer <
value
>
CALCulate:LIMit:LOWer <
value
>
–
The
STATus:QUEStionable:CONDition
register will indicate when an upper
or lower limit has been exceeded failing either a HI or LO limit test. Use the
STAT:QUES[:EVEN]?
command to query the status questionable register
and determine what failure occurred. Sending this command also clears
the questionable data register (or send a Clear Status
*CLS
command to
clear the register before testing begins).