clxxii
Keysight E1412A Digital Multimeter User Guide
Multimeter Command Reference
[SENSe:]
Example
Query the Autorange Mode
RES:RANG:AUTO OFF
Disable autorange.
RES:RANG:AUTO?
Query multimeter to return autorange mode.
enter statement
Enter value into computer.
RESistance:RESolution
[SENSe:]RESistance:RESolution <
resolution
>
selects the resolution for 2-wire
resistance measurements.
Parameters
Comments
–
MINimum
selects the best resolution (the smallest value) for the selected
range.
MAXimum
selects the worst resolution (the largest value) for the
selected range.
–
You must select a range using
RESistance:RANGe
before specifying
resolution. Also, only specify a resolution when making measurements on a
fixed range. Otherwise, the resolution will change to correspond with the
range selected during autoranging.
–
If autoranging is required, set the resolution using the
MIN
or
MAX
parameters.
–
If necessary to achieve the specified resolution, the multimeter will
increase the integration time as needed. This command overrides the
results of previously executed
RESistance:NPLC
or
RESistance:APERture
command (the last command executed has priority).
–
The
RES:RESolution
command overrides the resolution setting from a
previous
CONFigure:RESistanc
e command.
–
Related Commands:
CONFigur
e,
RESistance:NPLC
–
*RST
Condition: Based on the
*RST
values for the
RESistance:NPLC
command.
Example
Change the Resolution
Parameter Name
Parameter Type
Range of Values
Default
Units
<
resolution
>
numeric
resolution
|MIN|MAX
ohms
Содержание E1412A
Страница 1: ...Keysight E1412A 6 1 2 Digit Multimeter User Manual and SCPI Programming Guide 75000 Series C ...
Страница 2: ......
Страница 6: ...vi ...
Страница 18: ...xviii Keysight E1412A Multimeter User Guide ...
Страница 208: ...ccviii Keysight E1412A Digital Multimeter User Guide Multimeter Command Reference SCPI Command Quick Reference ...
Страница 242: ......