Keysight E1412A Digital Multimeter User Guide
clvii
[SENSe:]
Multimeter Command Reference
Comments
–
To select a standard measurement range, specify range as the input
signal’s maximum expected voltage. The multimeter then selects the
correct range.
–
MIN
selects the minimum range available with the
VOLTage :RANGe
command: 100mV.
MAX
selects the maximum range available: 300V.
–
Specifying a fixed range disables the autorange mode set by the
FREQ:VOLT:RANG:AUTO
command.
–
The
FREQ:VOLT:RANG
command overrides the range setting from a
previous
CONFigure:FREQuenc
y command.
–
*RST
Condition:
FREQ:VOLT:RANG 10
Example
Set the Voltage Range for Frequency Measurements to 100V
FREQ:VOLT:RANG 100
Voltage range for frequency
measurements is 100V.
FREQuency:VOLTage:RANGe?
[SENSe:]FREQuency:VOLTage:RANGe? [MIN|MAX]
returns one of the following
numbers to the output buffer: 0.1, 1, 10, 100 or 300.
–
The present voltage range selected if
MIN
or
MAX
is not specified.
–
The minimum voltage range available (100mV) if
MIN
is specified.
–
The maximum voltage range available (300V) if
MAX
is specified.
Example
Query the Measurement Range
FREQ:VOLT:RANG 10
Select 10 V range.
FREQ:VOLT:RANG?
Query the present range.
enter statement
Enter value into computer.
FREQuency:VOLTage:RANGe:AUTO
[SENSe:]FREQuency:VOLTage:RANGe:AUTO <
mode
>
enables or disables the
autorange function for the signal level of frequency measurements.
<
range
>
numeric
100mV|1V|10V|100V|300V|
MIN|MAX
volts