Keysight E1412A Digital Multimeter User Guide
cxlix
[SENSe:]
Multimeter Command Reference
–
The maximum current resolution available if
MAX
is specified.
Example
Query the AC Current Measurement Range
CURR:AC:RES 1E-4
Select 100
A resolution
CURR:AC:RES?
Query multimeter to return the present resolution.
enter statement
Enter value into computer.
CURRent[:DC]:APERture
[SENSe:]CURRent[:DC]:APERture <
number
>
sets the integration time in
seconds for dc current measurements. Values are rounded up to the nearest
aperture time shown in the following table.
Parameters
Comments
–
MIN
sets the aperture time to 0.333 ms.
MAX
sets the aperture time to
1.66667 seconds (60Hz) or 2 seconds (50Hz).
–
Setting the aperture time also sets the integration time in power line cycles
(PLCs) and the resolution. For example, an aperture time of 16.7ms (60Hz
line frequency) sets an integration time of 1 PLC. The corresponding
resolution depends on the function and range you select.
–
The
CURR:DC:APER
command overrides the results of previously executed
CURR:DC:NPLC
and
CURR:DC:RES
commands. The last command
executed has priority.
–
The greater the aperture time, the greater the normal mode rejection (and
the lower the reading rate).
–
Related Commands:
CALibration:LFRrequency
–
*RST
Condition:
CURR:DC:APER 0.166667
seconds (60Hz) or
CURR:DC:APER 0.20000
(50Hz)
Example
Set an Aperture Time of 16.7ms
CURR:APER 16.7E-03
Aperture time is 16.7 ms.
Parameter Name Parameter Type
Range of Values
Default
Units
<
number
>
numeric
0.333ms|3.33ms|16.7ms|
167ms|1.67s|MIN|MAX
seconds