Keysight E1412A Digital Multimeter User Guide
clxix
[SENSe:]
Multimeter Command Reference
Comments
–
MINimum
selects 0.02 PLCs.
MAXimum
selects 100 PLCs. Setting the
integration time in power line cycles (PLCs) also sets the integration time
and the resolution. For example, 10 PLCs sets an aperture time of 167ms
(60Hz line frequency) or 200ms (50Hz). The corresponding resolution
depends on the function and range you select.
–
The
RES:NPLC
command overrides the results of a previously executed
RESistance:RESolution
or
RESistance:APERture
command (the last
command executed has priority).
–
The greater the number of PLCs, the greater the normal mode rejection
(and the lower the reading rate).
–
Only the 1 PLC, 10 PLC and 100 PLC settings provide normal mode
rejection of 50Hz or 60Hz power line related noise.
–
*RST
Condition: 10 PLC
Example
Set the Integration Time in PLCs
RES:NPLC 100
Integration time is 100 PLCs.
RESistance:NPLC?
[SENSe:]RESistance:NPLC? [MIN|MAX]
returns one of the following numbers to
the output buffer:
–
The present integration time in PLCs if
MINimum
or
MAXimum
is not
specified.
–
The minimum integration time available (0.02) if
MIN
is specified.
–
The maximum integration time available (100) if
MAX
is specified.
Example
Query the Integration Time
RES:NPLC 100
Integration time is 100 PLCs.
RES:NPLC?
Query multimeter to return integration time.
enter statement
Enter value into computer.
RESistance:RANGe
[SENSe:]RESistance:RANGe <
range
>
selects the range for 2-wire resistance
measurements.