clxviii
Keysight E1412A Digital Multimeter User Guide
Multimeter Command Reference
[SENSe:]
–
Related Commands:
CALibration:LFRrequency
–
*RST
Condition:
RES:APER 0.166667
seconds (60Hz) or
RES:APER 0.20000
(50Hz)
Example
Set an Aperture Time of 16.7ms
RES:APER 16.7E-03
Aperture time is 16.7ms.
RESistance:APERture?
[SENSe:]RESistance:APERture? [MIN|MAX]
returns one of the following
numbers to the output buffer.
–
The present aperture time in seconds if
MIN
or
MAX
is not specified.
–
The minimum aperture time available (.333 ms) if
MIN
is specified.
–
The maximum aperture time available (1.67s @ 60Hz; 2s @ 50Hz) if
MAX
is specified.
Example
Query the Aperture Time
RES:APER 167E-03
Aperture time is 167ms.
RES:APER?
Query multimeter to return aperture time.
enter statement
Enter value into computer.
RESistance:NPLC
[SENSe:]RESistance:NPLCycles <
number
>
sets the integration time in power
line cycles (PLCs). The NPLC is set to a value from the range of values that can
accommodate the
<
number
>
you specify. For example, specifying
11
sets the
NPLC to 100.
Parameters
Parameter Name Parameter Type
Range of Values
Default
Units
<
number
>
numeric
0.02|0.2|1|10|100|MIN|MAX
PLCs