Keysight E1412A Digital Multimeter User Guide
Multimeter Command Reference
[SENSe:]
“
RES
”
“
VOLT:AC
”
“
VOLT
”
“
VOLT:RAT
”
Example
Query the Measurement Function
FUNC “FRES”
Function: 4-wire ohms.
FUNC?
Query multimeter to return selected function.
enter statement
Enter quoted string into computer.
CURRent:AC:RANGe
[SENSe:]CURRent:AC:RANGe <
range
>
selects the range for AC current
measurements.
Parameters
Comments
–
To select a standard measurement range, specify range as the input
signal’s maximum expected current. The multimeter then selects the
correct range.
–
MIN
selects the minimum range available with the
CURRent:AC:RANGe
command: 1A.
MAX
selects the maximum range available: 3A.
–
You must select a range using
CURRent:AC:RANGe
before specifying
resolution.
–
Specifying a fixed range disables the autorange mode set by the
CURR:AC:RANG:AUTO
command.
–
The
CURR:AC:RANG
command overrides the range setting from a previous
CONFigure
command on the same function.
–
*RST
Condition:
CURR:AC:RANG 1
CURRent:AC:RANGe?
[SENSe:]CURRent:AC:RANGe? [MIN|MAX]
returns one of the following numbers
to the output buffer:
Parameter Name
Parameter Type
Range of Values
Default Units
<
range
>
numeric
1A|3A|MIN|MAX
amps