
Keysight E1412A Digital Multimeter User Guide
clxvii
[SENSe:]
Multimeter Command Reference
Example
Disable Autoranging
PER:VOLT:RANG:AUTO OFF
Disable autorange.
PERiod:VOLTage:RANGe:AUTO?
[SENSe:]PERiod:VOLTage:RANGe:AUTO?
returns a number to show whether
the autorange mode is enabled or disabled: “
1
” =
ON
, “
0
” =
OFF
. The number is
sent to the output buffer.
Example
Query the Autorange Mode
PER:VOLT:RANG:AUTO OFF
Disable autorange.
PER:VOLT:RANG:AUTO?
Query multimeter to return autorange mode.
enter statement
Enter value into computer.
RESistance:APERture
[SENSe:]RESistance:APERture <
number
>
sets the integration time in seconds
for 2-wire resistance measurements. Values are rounded up to the nearest
aperture time shown in the following table.
Parameters
Comments
–
MIN
sets the aperture time to 0.333ms.
MAX
sets the aperture time to
1.66667 seconds (60Hz) or 2 seconds (50Hz).
–
Setting the aperture time also sets the integration time in power line cycles
(PLCs) and the resolution. For example, an aperture time of 16.7ms (60Hz
line frequency) sets an integration time of 1 PLC. The corresponding
resolution depends on the function and range you select.
–
The
RES:APER
command overrides the results of previously executed
RES:NPL
C and
RESistance:RESolution
commands. The last command
executed has priority.
–
The greater the aperture time, the greater the normal mode rejection (and
the lower the reading rate).
Parameter Name Parameter Type
Range of Values
Default
Units
<
number
>
numeric
0.333ms|3.33ms|16.7ms|
167ms|1.67s|MIN|MAX
seconds
Содержание E1412A
Страница 1: ...Keysight E1412A 6 1 2 Digit Multimeter User Manual and SCPI Programming Guide 75000 Series C ...
Страница 2: ......
Страница 6: ...vi ...
Страница 18: ...xviii Keysight E1412A Multimeter User Guide ...
Страница 208: ...ccviii Keysight E1412A Digital Multimeter User Guide Multimeter Command Reference SCPI Command Quick Reference ...
Страница 242: ......