Keysight E1412A Digital Multimeter User Guide
clxv
[SENSe:]
Multimeter Command Reference
Example
Set the Aperture Time
PER:APER 1E-2
Aperture time is 10 ms.
PERiod:APERture?
[SENSe:]PERiod:APERture? [MIN|MAX]
returns one of the following numbers to
the output buffer:
–
The present aperture time in seconds if MIN or MAX is not specified.
–
The minimum aperture time available (10 ms) if MIN is specified.
–
The maximum aperture time available (100 ms) if MAX is specified.
Example
Query the Aperture Time
PER:APER MIN
Aperture time is 10ms.
PER:APER?
Query multimeter to return aperture time.
enter statement
Enter value into computer.
PERiod:VOLTage:RANGe
[SENSe:]PERiod:VOLTage:RANGe <
range
>
selects the voltage range for the
signal level of period measurements.
Parameters
Comments
–
To select a standard measurement range, specify range as the input
signal’s maximum expected voltage. The multimeter then selects the
correct range.
–
MIN
selects the minimum range available with the
PERiod:VOLTage:RANGe
command: 100mV.
MAX
selects the maximum
range available: 300V.
–
Specifying a fixed range disables the autorange mode set by the
PER:VOLT:RANG:AUTO
command.
–
*RST
Condition:
PER:VOLT:RANG 10
Parameter Name Parameter Type
Range of Values
Default
Units
<
range
>
numeric
100mV|1V|10V|100V|300V|
MIN|MAX
volts