Keysight E1412A Digital Multimeter User Guide
cxci
TRIGger
Multimeter Command Reference
Example
Set the Trigger Count
CONF:VOLT:DC
Function: DC voltage.
TRIG:SOUR EXT
Trigger source is “Trig” BNC on multimeter
front panel.
TRIG:COUN 10
Multimeter will accept 10 external triggers
(one measurement is taken with each trigger).
READ?
Place multimeter in wait-for-trigger state; make
measurement when external trigger is received;
send readings to output buffer.
enter statement
Enter readings into computer.
:COUNt?
TRIGger:COUNt? [MIN|MAX]
returns one of the following numbers to the output
buffer:
–
The present trigger count (1 through 50,000) if
MIN
or
MAX
are not
specified.
–
The minimum trigger count available (1) if
MIN
is specified.
–
The maximum trigger count available (50,000) if
MAX
is specified.
Example
Query the Trigger Count
TRIG:COUN 10
Multimeter will accept 10 triggers.
TRIG:COUN?
Query multimeter to return trigger count.
enter statement
Enter value into computer.
:DELay
TRIGger:DELay <
seconds
>
sets the delay time between receipt of the trigger
and the start of the measurement. NOTE: This delay also occurs between each
sample when
SAMP:COUN > 1
. See triggering process diagram.
Parameters
Parameter Name
Parameter Type
Range of Values
Default
Units
<
seconds
>
numeric
0 through 3600|MIN|MAX
seconds