Keysight E1412A Digital Multimeter User Guide
cxli
SAMPle
Multimeter Command Reference
SAMPle
The
SAMPle
command subsystem operates with the TRIGger command
subsystem. The
SAMPle
subsystem designates the number of readings (count)
made for each trigger signal received.
Subsystem Syntax
SAMPle
:COUNt <
number
>|MIN|MAX
:COUNt? [MIN|MAX]
:COUNt
SAMPle:COUNt <
number
>|MIN|MAX
designates the number of readings per
trigger.
Parameters
Comments
–
MINimum
sets 1 reading per trigger.
MAXimum
sets 50,000 readings per
trigger.
–
If
MAX
or
50,000
is specified for
number
, the command executes without
error. When an
INIT
is executed requiring readings to be stored in internal
memory, an “Insufficient memory” error is generated to show that the
number of readings exceeds the memory available. However, you can
execute
READ?
which returns the readings to the output buffer and does
not use internal memory.
–
A number >50,000 returns Error -222, “Data out of range”.
–
CONFigure
and
MEASure
set the sample count to 1.
–
*RST Condition:
SAMP:COUN 1
Example
Set the Sample Count
CONF:VOLT:DC
Function: DC voltage.
TRIG:SOUR EXT
Trigger source is external BNC on
multimeter front panel.
Parameter Name Parameter Type
Range of Values
Default
Units
<
number
>
numeric
1 through 50,000|MIN|MAX
none