Keysight E1412A Digital Multimer User Guide
ccxxxvii
General Guidelines for Increasing Measurement Speed
Measurement Speed and Accuracy Trade-offs
The
EXTernal TRIGger
input can be used to start a scan from an external signal.
The E1406 Command Module “Clk out” is a convenient source for an external
signal. A potential problem exists whereby an external trigger arrives before the
multimeter is ready to start a new scan causing the trigger to be missed and no
error message generated. Send a
*OPC?
command to the multimeter after
setting up the multimeter and prior to initiating the switch module to eliminate
this problem.
Set Autozero to ONCE or OFF
Autozero causes the A/D to alternately measure its internal zero and the external
signal. Autozero improves reading accuracy; however, it reduces reading speed
by ½.
CAL:ZERO:AUTO ON
----
The zero will be measured before each
measurement.
CAL:ZERO:AUTO OFF
----
No new zero readings will be made.
CAL:ZERO:AUTO ONCE
----
Does one Autozero operation when the
command is received and also sets the mode
to autozero OFF.
The zero may vary over time, especially as the room temperature varies.
Noticeable changes can be expected over many minutes or hours. However, over
a few seconds the changes should be very small.
Turn Autorange OFF
Turning autorange
OFF
makes the E1412A take all measurements on a fixed
range which results in fast and predictable measurement times.
Autorange is turned
OFF
when a numeric value or
MIN
|
MAX
is specified for the
<
range
> parameter of the
CONFigure
,
MEASure
,
[SENSe:]RESistance:RANGe
,
[SENSe:]VOLTage:RANGe
, or
[SENSe:]VOLTage:AC:RANGe
command.
Autorange is directly controlled by the
[SENSe:]VOLTage:RANGe:AUTO ON|OFF
,
[SENSe:]VOLTage:AC:RANGe:AUTO ON|OFF
, or
[SENSe:]RESistance:RANGe:AUTO ON|OFF
command.
You can query the E1412A autorange status for a particular function by the
following commands:
[SENSe:]CURRent[:DC]:RANGe:AUTO?
[SENSe:]CURRent:AC:RANGe:AUTO?
[SENSe:]FREQuency:VOLTage:RANGe:AUTO?
[SENSe:]FRESistance:RANGe:AUTO?
[SENSe:]PERiod:VOLTage:RANGe:AUTO?
[SENSe:]RESistance:RANGe:AUTO?
[SENSe:]VOLTage[:DC]:RANGe:AUTO?
[SENSe:]VOLTage:AC:RANGe:AUTO?
[SENSe:]RESistance:RANGe:AUTO?
Содержание E1412A