Keysight N5290/1A Service Guide 4-7
N5290/91A Troubleshooting
N5292A Test Controller Troubleshooting
d.
Ensure the TSMB DC Supply indicator LEDs are on. If not, suspect
power supply (0950-0516) or front panel Standby switch. Using a
DVM measure the power supply terminal connections. Refer to the
label on the power supply for voltages on the terminals screws.
Refer to
“Verifying the A1 power supply” on page 4-17
.
e.
If the voltages on the power supply voltages are
not
correct:
i.
Turn off the power switch.
ii.
Then lift the TSMB.
iii.
If the DC Supply LEDs were on, disconnect bias cables (J60 &
J61) and amp cables (J10, 11, 12 and 14) to the TSMB and
re-install the board.
iv.
If the indicator LEDs are on, suspect the A3 LODA.
the A2 LOPA, A3 LODA, and A4 Phase Compensation Boards”
f.
If the fan is not working and all +3.3V are OK and the DC Supply
LEDs are on, replace fan. Refer to TP20, F6, and
,
, and
Содержание N5222B
Страница 3: ...3 ...
Страница 4: ...4 ...
Страница 10: ...6 Keysight N5292A User s and Service Guide Contents ...
Страница 28: ...1 18 Keysight N5290 1A Service Guide Safety and Regulatory Information Site Preparation 1 ...
Страница 86: ...3 50 Keysight N5290 1A Service Guide System Tests and Adjustments Adjustments 3 ...
Страница 128: ...5 10 Keysight N5290 1A Service Guide Theory of Operation Theory of Operation 5 ...
Страница 185: ...Keysight N5290 1A Service Guide 7 9 Repair and Replacement Procedures Removing the Cover Figure 7 3 Outer Cover Removal ...
Страница 214: ...7 38 Keysight N5290 1A Service Guide Repair and Replacement Procedures Post Repair Procedures 7 ...
Страница 224: ...8 10 Keysight N5290 1A Service Guide General Purpose Maintenance Procedures Correction Constants 8 ...
Страница 225: ...Installation Note Xxxxx xxxxx 3 ...