Keysight N5290/1A Service Guide 3-3
System Tests and Adjustments
Before You Begin
Before You Begin
Complete the Installation procedures in Chapter 3, "System Installation" of the
N5292-90002 N5290/1/A Microwave Network Analyzer System Installation
Guide to assemble and configure the system. After you press the Preeset
button, the PNA should display a frequency range of 900 Hz to 110 GHz
(N5290A) or 900 Hz to 120 GHz (N5291A). Refer to
https://literature.cdn.keysight.com/litweb/pdf/N5292-90002.pdf
.
Before checking, verifying, or adjusting the analyzer, refer to the following
paragraphs to:
— make sure the operating environment is within its requirements
— make sure that proper electrostatic discharge (ESD) protection is provided
— make sure the analyzer has warmed up properly to achieve system stability
— review the principles of connector care
Verify the Operating Environment
Due to their operating specifications, the verification and calibration kit devices
determine the limits of your operating environment conditions. Open the
calibration and verification kits and place all the devices on top of the foam
inserts so they will reach room temperature. As the device dimensions change
with temperature, their electrical characteristics change as well.
It is necessary to keep the environmental levels within the following limits:
— Temperature: +23
°
C ± 3
°
C (Error-corrected temperature range)
Once the measurement calibration has been done, the ambient
temperature must be maintained to within
±
1
°
C of the calibration
temperature.
— Humidity: 0% to 95% at 40
°
C maximum, non-condensing
—
Altitude: 0 to 4,600 meters (
≈
15,000 feet.)
IMPORTANT!
Before you begin any testing on a System or instrument,
always inspect the connectors on the instrument, the calibration kit,
cables and adapters.
to review where the
calibration stickers should be placed on the PNA.
IMPORTANT!
Before you begin any testing on a System or instrument,
always inspect the connectors on the instrument, the calibration kit,
cables and adapters.
Содержание N5222B
Страница 3: ...3 ...
Страница 4: ...4 ...
Страница 10: ...6 Keysight N5292A User s and Service Guide Contents ...
Страница 28: ...1 18 Keysight N5290 1A Service Guide Safety and Regulatory Information Site Preparation 1 ...
Страница 86: ...3 50 Keysight N5290 1A Service Guide System Tests and Adjustments Adjustments 3 ...
Страница 128: ...5 10 Keysight N5290 1A Service Guide Theory of Operation Theory of Operation 5 ...
Страница 185: ...Keysight N5290 1A Service Guide 7 9 Repair and Replacement Procedures Removing the Cover Figure 7 3 Outer Cover Removal ...
Страница 214: ...7 38 Keysight N5290 1A Service Guide Repair and Replacement Procedures Post Repair Procedures 7 ...
Страница 224: ...8 10 Keysight N5290 1A Service Guide General Purpose Maintenance Procedures Correction Constants 8 ...
Страница 225: ...Installation Note Xxxxx xxxxx 3 ...