3-40 Keysight N5290/1A Service Guide
System Tests and Adjustments
N5292A Functional Verification Procedure
3-
Test Set Connector and Cable Verification
Equipment required:
—
— S-Parameters Fixture - Refer to
— N5292A Test Set and Cable verification Launch Test application (
PNA Cal
)
(
)
through
Procedure:
— On the PNA: Using test cables, connect ECal module between ports 1 and
2. Refer to
4.
Perform base line 2-port calibration using Launch Test application (
PNA
CAL
) and ECal module. Follow the prompts.
— On the N5292A Test Set: Connect the S-Parameters fixture to the port
under test. Refer to
5.
Using the Launch Test application (
Insertion Loss/Match
), perform follow
prompts and compare results to the typical plots shown in
through
Figure 3-31
N5292A Front Panel Integrated Port Connector
When a defective cable or connector has been replaced and the test set
has been repaired, the following tests are required:
—
Test set and N5290A/91 A system calibration
— Insertion Loss/Match-S-parameters (Refer to
through
.)
—
LO Verify (Run after Insertion Loss/Match are verified.)
IMPORTANT!
LO power verification is only required on the port that the
connector or and cabling was replaced.
Содержание N5222B
Страница 3: ...3 ...
Страница 4: ...4 ...
Страница 10: ...6 Keysight N5292A User s and Service Guide Contents ...
Страница 28: ...1 18 Keysight N5290 1A Service Guide Safety and Regulatory Information Site Preparation 1 ...
Страница 86: ...3 50 Keysight N5290 1A Service Guide System Tests and Adjustments Adjustments 3 ...
Страница 128: ...5 10 Keysight N5290 1A Service Guide Theory of Operation Theory of Operation 5 ...
Страница 185: ...Keysight N5290 1A Service Guide 7 9 Repair and Replacement Procedures Removing the Cover Figure 7 3 Outer Cover Removal ...
Страница 214: ...7 38 Keysight N5290 1A Service Guide Repair and Replacement Procedures Post Repair Procedures 7 ...
Страница 224: ...8 10 Keysight N5290 1A Service Guide General Purpose Maintenance Procedures Correction Constants 8 ...
Страница 225: ...Installation Note Xxxxx xxxxx 3 ...