3-44 Keysight N5290/1A Service Guide
System Tests and Adjustments
If the N5290/1A Operator’s Check Fails
3-
If the N5290/1A Operator’s Check Fails
1.
Clean the test ports, shorts, and adapters. Torque to specification. Repeat
the check.
2.
Run the adjustment tests and repeat the Operator’s Check. Refer
.
3.
If the check still fails, suspect a faulty component. Refer to
to begin troubleshooting to determine the
faulty component.
4.
Also, refer to
Содержание N5222B
Страница 3: ...3 ...
Страница 4: ...4 ...
Страница 10: ...6 Keysight N5292A User s and Service Guide Contents ...
Страница 28: ...1 18 Keysight N5290 1A Service Guide Safety and Regulatory Information Site Preparation 1 ...
Страница 86: ...3 50 Keysight N5290 1A Service Guide System Tests and Adjustments Adjustments 3 ...
Страница 128: ...5 10 Keysight N5290 1A Service Guide Theory of Operation Theory of Operation 5 ...
Страница 185: ...Keysight N5290 1A Service Guide 7 9 Repair and Replacement Procedures Removing the Cover Figure 7 3 Outer Cover Removal ...
Страница 214: ...7 38 Keysight N5290 1A Service Guide Repair and Replacement Procedures Post Repair Procedures 7 ...
Страница 224: ...8 10 Keysight N5290 1A Service Guide General Purpose Maintenance Procedures Correction Constants 8 ...
Страница 225: ...Installation Note Xxxxx xxxxx 3 ...