Keysight N5290/1A Service Guide 8-5
General Purpose Maintenance Procedures
Keysight License Manager
Enabling or Removing Options
There are two types of options:
— Hardware: Hardware options involve adding additional hardware to the
analyzer. After the proper hardware has been installed in the analyzer, the
appropriate license can be installed using the Keysight License Manager.
It is necessary to re-enable all installed hardware options, if the test set
mother-board is replaced.
— Software: Software options add features or functionality to the analyzer
without the need for additional hardware. These options are enabled by
installing the appropriate licenses using the Keysight License Manager
(KLM). Refer to
http://literature.cdn.keysight.com/litweb/pdf/N5242-90024.pdf
To enable or remove an option:
1.
Start the Keysight Software Manager. Refer to
.
2.
In the Keysight License Manager window that opens, in the left hand
column, press
local host
>
Local Licenses
. A list of available options,
similar to the list below, will appear. Refer to
S93015A - Dynamic uncertainty for S-parameters
S93025A - Basic pulsed-RF measurements
029 - Noise figure measurements
S93086A - Gain Compression Application
S93087A- IMD Measurements
029 - Full Noise Measurements
219 - Src/Rcvr Atten & Bias Ts 2-Port
419 - Src/Rcvr Atten & Bias Ts 4-Port
224 - 2nd Src w/Combiner & Switches
423 - Combiner & Switches
3.
Press and hold (or right click with a mouse) the option that you wish to
either enable or remove, and then click
Install
or
Delete
, whichever is
appropriate.
Some applications require a license key file that is provided by Keysight. If
you do not have the required license key file, contact Keysight for
“Contacting Keysight” on page 2-7
.
Содержание N5222B
Страница 3: ...3 ...
Страница 4: ...4 ...
Страница 10: ...6 Keysight N5292A User s and Service Guide Contents ...
Страница 28: ...1 18 Keysight N5290 1A Service Guide Safety and Regulatory Information Site Preparation 1 ...
Страница 86: ...3 50 Keysight N5290 1A Service Guide System Tests and Adjustments Adjustments 3 ...
Страница 128: ...5 10 Keysight N5290 1A Service Guide Theory of Operation Theory of Operation 5 ...
Страница 185: ...Keysight N5290 1A Service Guide 7 9 Repair and Replacement Procedures Removing the Cover Figure 7 3 Outer Cover Removal ...
Страница 214: ...7 38 Keysight N5290 1A Service Guide Repair and Replacement Procedures Post Repair Procedures 7 ...
Страница 224: ...8 10 Keysight N5290 1A Service Guide General Purpose Maintenance Procedures Correction Constants 8 ...
Страница 225: ...Installation Note Xxxxx xxxxx 3 ...