Keysight N5290/1A Service Guide 3-31
System Tests and Adjustments
N5292A Functional Verification Procedure
Figure 3-19
Screwdriver and TSMB with LEDs
S-Parameters (Insertion Loss/Match)
Equipment required:
— 26.5 GHz PNA-B with Option 020
— ECal or Calibration kit with a minimum 26.5 GHz frequency
— RF Fixture (ET57259-1, N5292-60128)
— S-Parameters fixture (ET57259-2, N5292-60129)
— PC with Launch Test application loaded (
)
1.
N5292A Test Set Calibration (using Launch Test application)
S-parameter Testing: Connect ECal module between ports 1 and 2. Using
an ECal or calibration kit and
Pna Cal
button in the Launch Test
application, do a 2-port cal to verify cabling and PNA integrity.
2.
When 2-port calibration has been completed, proceed to the test set
insertion loss/transmission measurements done after the N5290/91 A
This section assumes you have already downloaded the Launch Test
application to your PC. If not, refer to
“PNA Calibration (Pna Cal) and
Loading the N5292A Test Set Functional Verification Software (Launch
Содержание N5222B
Страница 3: ...3 ...
Страница 4: ...4 ...
Страница 10: ...6 Keysight N5292A User s and Service Guide Contents ...
Страница 28: ...1 18 Keysight N5290 1A Service Guide Safety and Regulatory Information Site Preparation 1 ...
Страница 86: ...3 50 Keysight N5290 1A Service Guide System Tests and Adjustments Adjustments 3 ...
Страница 128: ...5 10 Keysight N5290 1A Service Guide Theory of Operation Theory of Operation 5 ...
Страница 185: ...Keysight N5290 1A Service Guide 7 9 Repair and Replacement Procedures Removing the Cover Figure 7 3 Outer Cover Removal ...
Страница 214: ...7 38 Keysight N5290 1A Service Guide Repair and Replacement Procedures Post Repair Procedures 7 ...
Страница 224: ...8 10 Keysight N5290 1A Service Guide General Purpose Maintenance Procedures Correction Constants 8 ...
Страница 225: ...Installation Note Xxxxx xxxxx 3 ...