:SBUS<n>:SENT:TRIGger
Command Syntax
:SBUS<n>:SENT:TRIGger <mode>
<mode> ::= {SFCMessage | SSCMessage | FCData | SCMid | SCData
| TOLerror | FCCerror | SCCerror | CRCerror | PPERror | SSPerror}
The :SBUS<n>:SENT:TRIGger command specifies the SENT trigger mode:
• SFCMessage — triggers on the start of any Fast Channel Message (after 56
Synchronization/Calibration ticks).
• SSCMessage — trigger on the start of any Slow Channel Message.
• FCData — triggers on a Fast Channel Message when the Status &
Communication nibble and the data nibbles match the values entered using
additional softkeys.
• SCMid — triggers when a Slow Channel Message ID matches the value entered
using additional softkeys.
• SCData — triggers when a Slow Channel Message ID and Data field both match
the values entered using additional softkeys.
• TOLerror — triggers when the sync pulse width varies from the nominal value by
greater than the entered percentage.
• FCCerror — triggers on any Fast Channel Message CRC error.
• SCCerror — triggers on any Slow Channel Message CRC error.
• CRCerror — triggers on any CRC error, Fast or Slow.
• PPERror — triggers if a nibble is either too wide or too narrow (for example, data
nibble < 12 (11.5) or > 27 (27.5) ticks wide). Sync, S&C, data, or checksum
pulse periods are checked.
• SSPerror — triggers on a sync pulse whose width varies from the previous sync
pulse's width by greater than 1/64 (1.5625%, as defined in the SENT
specification).
Query Syntax
:SBUS<n>:SENT:TRIGger?
The :SBUS<n>:SENT:TRIGger? query returns the trigger mode setting.
Return Format
<mode><NL>
<mode> ::= {SFCM | SSCM | FCD | SCM | SCD | TOL | FCC | SCC | CRC
| PPER | SSP}
See Also
•
•
•
•
Содержание InfiniiVision M9241A
Страница 1: ...Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide ...
Страница 52: ...52 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 2 Setting Up ...
Страница 178: ...178 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 5 Commands Quick Reference ...
Страница 246: ...246 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 7 Root Commands ...
Страница 268: ...268 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 8 ACQuire Commands ...
Страница 280: ...280 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 9 CALibrate Commands ...
Страница 354: ...354 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 15 EXTernal Trigger Commands ...
Страница 412: ...412 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 17 FUNCtion m Commands ...
Страница 416: ...416 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 18 HCOPY Commands ...
Страница 442: ...442 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 20 MARKer Commands ...
Страница 588: ...588 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 23 MTESt Commands ...
Страница 692: ...692 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 24 POWer Commands ...
Страница 1062: ...1062 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 30 TIMebase Commands ...
Страница 1250: ...1250 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 33 WGEN w Commands ...
Страница 1306: ...1306 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 35 Obsolete and Discontinued Commands ...
Страница 1314: ...1314 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 36 Error Messages ...
Страница 1376: ...1376 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 39 More About Oscilloscope Commands ...