402
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
17
:FUNCtion<m> Commands
This function uses a Hilbert transform to get the real (in-phase, I) and
imaginary (quadrature, Q) parts of the input signal and then performs a
square root of the sum of the real and imaginary parts to get the
demodulated amplitude envelope waveform.
Filters operate on a single analog channel source or on a lower math function.
• Visualizations:
•
MAGNify — Operates on a single analog channel source or on a lower math
function.
•
MAXimum — This operator is like the MAXHold operator without the hold.
The maximum vertical values found at each horizontal bucket are used to
build a waveform.
•
MINimum — This operator is like the MINHold operator without the hold. The
minimum vertical values found at each horizontal bucket are used to build a
waveform.
•
PEAK — The PEAK operator is like the MAXimum operator minus the
MINimum operator. At each horizontal bucket, the minimum vertical values
found are subtracted from the maximum vertical values found to build a
waveform.
•
MAXHold — Operates on a single analog channel source or on a lower math
function. The Max Hold (or Max Envelope) operator records the maximum
vertical values found at each horizontal bucket across multiple analysis
cycles and uses those values to build a waveform.
•
MINHold — Operates on a single analog channel source or on a lower math
function. The Min Hold (or Min Envelope) operator records the minimum
vertical values found at each horizontal bucket across multiple analysis
cycles and uses those values to build a waveform.
•
TRENd — Measurement trend — Operates on a single analog channel source.
The TRENd:NMEasurement command selects the measurement whose trend
you want to measure.
Query Syntax
:FUNCtion<m>:OPERation?
The :FUNCtion<m>:OPERation? query returns the current operation for the
selected function.
Return Format
<operation><NL>
<operation> ::= {ADD | SUBT | MULT | DIV | INT | DIFF | FFT | FFTP
| SQRT | MAGN | ABS | SQU | LN | LOG | EXP | TEN | LOWP | HIGH
| AVER | SMO | ENV | LIN | MAX | MIN | PEAK | MAXH | MINH | TREN}
NOTE
If a math function is on (see :FUNCtion<m>:DISPlay), changing the operator will cause an
automatic vertical scaling of the new math function waveform. If you wait for the operation to
complete (with an *OPC? query for example), then query the math functions's scale, you can
see the vertical scaling that was automatically determined.
Содержание InfiniiVision M9241A
Страница 1: ...Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide ...
Страница 52: ...52 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 2 Setting Up ...
Страница 178: ...178 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 5 Commands Quick Reference ...
Страница 246: ...246 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 7 Root Commands ...
Страница 268: ...268 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 8 ACQuire Commands ...
Страница 280: ...280 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 9 CALibrate Commands ...
Страница 354: ...354 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 15 EXTernal Trigger Commands ...
Страница 412: ...412 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 17 FUNCtion m Commands ...
Страница 416: ...416 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 18 HCOPY Commands ...
Страница 442: ...442 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 20 MARKer Commands ...
Страница 588: ...588 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 23 MTESt Commands ...
Страница 692: ...692 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 24 POWer Commands ...
Страница 1062: ...1062 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 30 TIMebase Commands ...
Страница 1250: ...1250 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 33 WGEN w Commands ...
Страница 1306: ...1306 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 35 Obsolete and Discontinued Commands ...
Страница 1314: ...1314 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 36 Error Messages ...
Страница 1376: ...1376 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 39 More About Oscilloscope Commands ...