Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
38
Synchronizing Acquisitions
#
# How it works:
#
#
-
The :DIGitize command is a blocking command, and thus, all
#
other SCPI commands are blocked until :DIGitize is completely
#
done.
This includes any subsequent processing that is already
#
set up, such as math and measurements.
#
#
KEY POINT: However, :DIGitize does not prevent additional
#
commands from being sent to the queue or cause the remote
#
program to wait.
For example, if your program does something
#
like:
#
#
KsInfiniiVisionX.write(":DIGitize")
#
print("Signal acquired.\n")
#
#
The "Signal acquired" message will be written immediately
#
after the :DIGitize is sent, not after the acquisition and
#
processing is complete.
#
#
To pause the program until the :DIGitize is complete, you must
#
wait for a query result after the :DIGitize.
For example, in
#
this case:
#
#
query_result = KsInfiniiVisionX.query(":DIGitize;*OPC?")
#
print("Signal acquired.\n")
#
#
The "Signal acquired" message will be written after the
#
acquisition and processing is complete.
The *OPC? query is
#
appended to :DIGitize with a semi-colon (;), which
#
essentially ties it to the same thread in the parser. It is
#
immediately dealt with once :DIGitize finishes and gives a "1"
#
back to the program (whether the program uses it or not),
#
allowing the program to move on.
#
# Other Notes:
#
#
-
If you DO NOT know when a trigger will occur, you should set a
#
very long timeout.
#
#
-
The timeout should be adjusted before and after the :DIGitize
#
operation, though this is not absolutely required.
#
#
-
A :DIGitize can be aborted with a device clear, which also
#
stops the oscilloscope:
#
KsInfiniiVisionX.clear()
#
#
-
:DIGItize disables the anti-aliasing feature (sample rate
#
dither) on all InfiniiVision and InfiniiVision X-Series
#
oscilloscopes.
#
#
-
:DIGitize temporarily blocks the front panel, and all front
#
panel presses are queued until :DIGitize is done.
So if you
#
change the vertical scale, it will not happen until the
#
acquisition is done.
#
#
The exception is that the Run/Stop button on the front panel
Содержание InfiniiVision M9241A
Страница 1: ...Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide ...
Страница 52: ...52 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 2 Setting Up ...
Страница 178: ...178 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 5 Commands Quick Reference ...
Страница 246: ...246 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 7 Root Commands ...
Страница 268: ...268 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 8 ACQuire Commands ...
Страница 280: ...280 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 9 CALibrate Commands ...
Страница 354: ...354 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 15 EXTernal Trigger Commands ...
Страница 412: ...412 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 17 FUNCtion m Commands ...
Страница 416: ...416 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 18 HCOPY Commands ...
Страница 442: ...442 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 20 MARKer Commands ...
Страница 588: ...588 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 23 MTESt Commands ...
Страница 692: ...692 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 24 POWer Commands ...
Страница 1062: ...1062 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 30 TIMebase Commands ...
Страница 1250: ...1250 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 33 WGEN w Commands ...
Страница 1306: ...1306 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 35 Obsolete and Discontinued Commands ...
Страница 1314: ...1314 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 36 Error Messages ...
Страница 1376: ...1376 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 39 More About Oscilloscope Commands ...