:ACQuire Commands
:ACQuire:TYPE
Command Syntax
:ACQuire:TYPE <type>
<type> ::= {NORMal | AVERage | HRESolution | PEAK}
The :ACQuire:TYPE command selects the type of data acquisition that is to take
place. The acquisition types are:
• NORMal — sets the oscilloscope in the normal mode.
• AVERage — sets the oscilloscope in the averaging mode. You can set the count
by sending the :ACQuire:COUNt command followed by the number of averages.
In this mode, the value for averages is an integer from 1 to 65536. The COUNt
value determines the number of averages that must be acquired.
The AVERage type is not available when in segmented memory mode
(:ACQuire:MODE SEGMented).
• HRESolution — sets the oscilloscope in the high-resolution mode (also known
as
smoothing
). This mode is used to reduce noise at slower sweep speeds
where the digitizer samples faster than needed to fill memory for the displayed
time range.
For example, if the digitizer samples at 200 MSa/s, but the effective sample
rate is 1 MSa/s (because of a slower sweep speed), only 1 out of every 200
samples needs to be stored. Instead of storing one sample (and throwing others
away), the 200 samples are averaged together to provide the value for one
display point. The slower the sweep speed, the greater the number of samples
that are averaged together for each display point.
• PEAK — sets the oscilloscope in the peak detect mode. In this mode,
:ACQuire:COUNt has no meaning.
The AVERage and HRESolution types can give you extra bits of vertical resolution.
See the
User's Guide
for an explanation. When getting waveform data acquired
using the AVERage and HRESolution types, be sure to use the WORD or ASCii
waveform data formats to get the extra bits of vertical resolution.
Query Syntax
:ACQuire:TYPE?
The :ACQuire:TYPE? query returns the current acquisition type.
Return Format
<acq_type><NL>
<acq_type> ::= {NORM | AVER | HRES | PEAK}
NOTE
The obsolete command ACQuire:TYPE:REALtime is functionally equivalent to sending
ACQuire:MODE RTIME; TYPE NORMal.
Содержание InfiniiVision M9241A
Страница 1: ...Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide ...
Страница 52: ...52 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 2 Setting Up ...
Страница 178: ...178 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 5 Commands Quick Reference ...
Страница 246: ...246 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 7 Root Commands ...
Страница 268: ...268 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 8 ACQuire Commands ...
Страница 280: ...280 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 9 CALibrate Commands ...
Страница 354: ...354 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 15 EXTernal Trigger Commands ...
Страница 412: ...412 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 17 FUNCtion m Commands ...
Страница 416: ...416 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 18 HCOPY Commands ...
Страница 442: ...442 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 20 MARKer Commands ...
Страница 588: ...588 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 23 MTESt Commands ...
Страница 692: ...692 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 24 POWer Commands ...
Страница 1062: ...1062 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 30 TIMebase Commands ...
Страница 1250: ...1250 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 33 WGEN w Commands ...
Страница 1306: ...1306 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 35 Obsolete and Discontinued Commands ...
Страница 1314: ...1314 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 36 Error Messages ...
Страница 1376: ...1376 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 39 More About Oscilloscope Commands ...