:TRIGger Commands
31
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
1129
:TRIGger:PXI:MODE
Command Syntax
:TRIGger:PXI:MODE <mode>
<mode> ::= {OFF | MASTer | SLAVe}
The :TRIGger:PXI:MODE command specifies the PXIe oscilloscope multi-module
triggering mode:
• OFF — This module does not participate in multi-module triggering and is a
standalone module.
• MASTer — This module is the master module.
The :TRIGger commands on the master module define the trigger condition.
After waiting for "armed" signals from each slave module, the master module
outputs the trigger signal on the selected trigger line.
• SLAVe — This module is a slave module.
Slave modules send an "armed" signal on their selected arm like and are
triggered by the selected PXIe trigger line.
It is important to identify the master slot number on each slave
(:TRIGger:PXI:MSLot) so that the proper set of calibration factors can be used
when aligning waveforms. Calibration factors are created and saved using the
KtScopeCal.exe application on the chassis controller.
Among all the oscilloscope modules to be synchronized, one module must be the
master and all other modules must be slaves.
Query Syntax
:TRIGger:PXI:MODE?
The :TRIGger:PXI:MODE? query returns the PXIe oscilloscope's multi-module
triggering mode.
Return Format
<mode><NL>
<mode> ::= {OFF | MAST | SLAV}
See Also
•
":TRIGger:PXI:MALine<n>:ENABle"
•
•
•
•
•
Содержание InfiniiVision M9241A
Страница 1: ...Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide ...
Страница 52: ...52 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 2 Setting Up ...
Страница 178: ...178 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 5 Commands Quick Reference ...
Страница 246: ...246 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 7 Root Commands ...
Страница 268: ...268 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 8 ACQuire Commands ...
Страница 280: ...280 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 9 CALibrate Commands ...
Страница 354: ...354 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 15 EXTernal Trigger Commands ...
Страница 412: ...412 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 17 FUNCtion m Commands ...
Страница 416: ...416 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 18 HCOPY Commands ...
Страница 442: ...442 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 20 MARKer Commands ...
Страница 588: ...588 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 23 MTESt Commands ...
Страница 692: ...692 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 24 POWer Commands ...
Страница 1062: ...1062 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 30 TIMebase Commands ...
Страница 1250: ...1250 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 33 WGEN w Commands ...
Страница 1306: ...1306 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 35 Obsolete and Discontinued Commands ...
Страница 1314: ...1314 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 36 Error Messages ...
Страница 1376: ...1376 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 39 More About Oscilloscope Commands ...