:MEASure Commands
21
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
481
:MEASure:FFT:ACPR
Command Syntax
:MEASure:FFT:ACPR <chan_width>,<chan_spacing>,<chan>[,<source>]
<chan_width> ::= of main range and sideband channels, Hz in NR3 format
<chan_spacing> ::= spacing between main range and sideband channels,
Hz in NR3 format
<chan> ::= {CENTer | HIGH<sb> | LOW<sb>}
<sb> ::= sideband 1 to 5
<source> ::= {FUNCtion<m> | MATH<m> | FFT} (must be an FFT waveform)
<m> ::= 1 to (# math functions) in NR1 format
The :MEASure:FFT:ACPR command installs an FFT analysis Adjacent Channel
Power Ratio (ACPR) measurement on screen.
Adjacent Channel Power Ratio (or channel leakage ratio) measures the ratio of the
power in the main frequency range to the power contained in one or more
sidebands.
The main range is specified by a channel width and a center frequency. The center
frequency used in the measurement is the one defined for the FFT function.
Sidebands (with the same width as the main range) exist above and below the
main range separated by the channel spacing width.
The sideband used for the measurement is selected with the <chan> parameter.
You can select the first through fifth sidebands above or below the main range
(HIGH1 through HIGH5 above and LOW1 through LOW5 below). The full sideband
must be in the graticule (on screen) to be measured. Otherwise, the measurement
results will be "Incomplete".
When this measurement is tracked with cursors, the cursors show the sideband
being measured.
Query Syntax
:MEASure:FFT:ACPR? <chan_width>,<chan_spacing>,<chan>[,<source>]
The :MEASure:FFT:ACPR? query returns the measured Adjacent Channel Power
Ratio (ACPR) value.
Return Format
<return_value><NL>
<return_value> ::= adjacent channel power ratio, dBV in NR3 format
See Also
•
•
•
Содержание InfiniiVision M9241A
Страница 1: ...Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide ...
Страница 52: ...52 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 2 Setting Up ...
Страница 178: ...178 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 5 Commands Quick Reference ...
Страница 246: ...246 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 7 Root Commands ...
Страница 268: ...268 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 8 ACQuire Commands ...
Страница 280: ...280 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 9 CALibrate Commands ...
Страница 354: ...354 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 15 EXTernal Trigger Commands ...
Страница 412: ...412 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 17 FUNCtion m Commands ...
Страница 416: ...416 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 18 HCOPY Commands ...
Страница 442: ...442 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 20 MARKer Commands ...
Страница 588: ...588 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 23 MTESt Commands ...
Страница 692: ...692 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 24 POWer Commands ...
Страница 1062: ...1062 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 30 TIMebase Commands ...
Страница 1250: ...1250 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 33 WGEN w Commands ...
Страница 1306: ...1306 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 35 Obsolete and Discontinued Commands ...
Страница 1314: ...1314 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 36 Error Messages ...
Страница 1376: ...1376 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 39 More About Oscilloscope Commands ...