Obsolete and Discontinued Commands
35
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
1297
:MTESt:RUMode
Command Syntax
:MTESt:RUMode {FORever | TIME,<seconds> | {WAVeforms,<wfm_count>}}
<seconds> ::= from 1 to 86400 in NR3 format
<wfm_count> ::= number of waveforms in NR1 format
from 1 to 1,000,000,000
The :MTESt:RUMode command determines the termination conditions for the
mask test. The choices are FORever, TIME, or WAVeforms.
• FORever — runs the Mask Test until the test is turned off.
• TIME — sets the amount of time in seconds that a mask test will run before it
terminates. The <seconds> parameter is a real number from 1 to 86400
seconds.
• WAVeforms — sets the maximum number of waveforms that are required before
the mask test terminates. The <wfm_count> parameter indicates the number of
waveforms that are to be acquired; it is an integer from 1 to 1,000,000,000.
Query Syntax
:MTESt:RUMode?
The :MTESt:RUMode? query returns the currently selected termination condition
and value.
Return Format
{FOR | TIME,<seconds> | {WAV,<wfm_count>}}<NL>
<seconds> ::= from 1 to 86400 in NR3 format
<wfm_count> ::= number of waveforms in NR1 format
from 1 to 1,000,000,000
See Also
•
"Introduction to :MTESt Commands"
•
NOTE
The :MTESt:RUMode command is obsolete and is provided for backward compatibility to
previous oscilloscopes. Use the :MTESt:RMODe command (see
) instead.
Содержание InfiniiVision M9241A
Страница 1: ...Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide ...
Страница 52: ...52 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 2 Setting Up ...
Страница 178: ...178 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 5 Commands Quick Reference ...
Страница 246: ...246 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 7 Root Commands ...
Страница 268: ...268 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 8 ACQuire Commands ...
Страница 280: ...280 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 9 CALibrate Commands ...
Страница 354: ...354 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 15 EXTernal Trigger Commands ...
Страница 412: ...412 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 17 FUNCtion m Commands ...
Страница 416: ...416 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 18 HCOPY Commands ...
Страница 442: ...442 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 20 MARKer Commands ...
Страница 588: ...588 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 23 MTESt Commands ...
Страница 692: ...692 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 24 POWer Commands ...
Страница 1062: ...1062 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 30 TIMebase Commands ...
Страница 1250: ...1250 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 33 WGEN w Commands ...
Страница 1306: ...1306 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 35 Obsolete and Discontinued Commands ...
Страница 1314: ...1314 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 36 Error Messages ...
Страница 1376: ...1376 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 39 More About Oscilloscope Commands ...