774
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
27
:SBUS<n> Commands
:SBUS<n>:CAN:TRIGger
Command Syntax
:SBUS<n>:CAN:TRIGger <condition>
<condition> ::= {SOF | EOF | IDData | DATA | FDData | IDRemote
| IDEither | ERRor | ACKerror | FORMerror | STUFferror | CRCerror
| SPECerror | ALLerrors | BRSBit | CRCDbit | EBActive | EBPassive
| OVERload | MESSage | MSIGnal | FDMSignal}
The :SBUS<n>:CAN:TRIGger command sets the CAN trigger on condition:
Condition
Front-panel name
Description
Filter by
ID*
SOF
SOF - Start of Frame
Triggers at the start bit for both data and
overload frames.
EOF
EOF - End of Frame
Triggers at the end of any frame.
X
IDEither
Frame ID
Triggers on any standard CAN (data or
remote) or CAN FD frame at the end of the
11- or 29-bit ID field.
IDData
Data Frame ID (non-FD)
Triggers on standard CAN data frames at
the end of the 11- or 29-bit ID field.
DATA
Data Frame ID and Data
(non-FD)
Triggers on any standard CAN data frame at
the end of the last data byte defined in the
trigger. The DLC of the packet must must
match the number of bytes specified.
FDData
Data Frame ID and Data
(FD)
triggers on CAN FD frames at the end of the
last data byte defined in the trigger. You can
trigger on up to 8 bytes of data anywhere
within the CAN FD data, which can be up to
64 bytes long.
IDRemote
Remote Frame ID
Triggers on standard CAN remote frames at
the end of the 11- or 29-bit ID field.
ERRor
Error Frame
Triggers after 6 consecutive 0s while in a
data frame, at the EOF.
X
ACKerror
Acknowledge Error
Triggers on the acknowledge bit if the
polarity is incorrect.
X
FORMerror Form Error
Triggers on reserved bit errors.
X
STUFferror Stuff Error
Triggers on 6 consecutive 1s or 6
consecutive 0s, while in a non-error or non
overload frame.
X
Содержание InfiniiVision M9241A
Страница 1: ...Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide ...
Страница 52: ...52 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 2 Setting Up ...
Страница 178: ...178 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 5 Commands Quick Reference ...
Страница 246: ...246 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 7 Root Commands ...
Страница 268: ...268 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 8 ACQuire Commands ...
Страница 280: ...280 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 9 CALibrate Commands ...
Страница 354: ...354 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 15 EXTernal Trigger Commands ...
Страница 412: ...412 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 17 FUNCtion m Commands ...
Страница 416: ...416 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 18 HCOPY Commands ...
Страница 442: ...442 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 20 MARKer Commands ...
Страница 588: ...588 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 23 MTESt Commands ...
Страница 692: ...692 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 24 POWer Commands ...
Страница 1062: ...1062 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 30 TIMebase Commands ...
Страница 1250: ...1250 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 33 WGEN w Commands ...
Страница 1306: ...1306 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 35 Obsolete and Discontinued Commands ...
Страница 1314: ...1314 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 36 Error Messages ...
Страница 1376: ...1376 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 39 More About Oscilloscope Commands ...