:POWer Commands
24
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
615
:POWer:DESKew
Command Syntax
:POWer:DESKew
The :POWer:DESKew command launches the auto deskew process on the
oscilloscope.
Before sending this command:
1
Demagnetize and zero-adjust the current probe.
Refer to the current probe's documentation for instructions on how to do this.
2
Make connections to the U1880A deskew fixture as described in the
oscilloscope's connection dialog or in the
Power Measurement Application
User's Guide
.
3
Make sure the voltage probe and current probe channels are specified
appropriately using the :POWer:SIGNals:SOURce:VOLTage1 and
:POWer:SIGNals:SOURce:CURRent1 commands.
The deskew values are saved in the oscilloscope until a factory default or secure
erase is performed. The next time you run the Power Application, you can use the
saved deskew values or perform the deskew again.
Generally, you need to perform the deskew again when part of the test setup
changes (for example, a different probe, different oscilloscope channel, etc.) or
when the ambient temperature has changed.
See Also
•
":POWer:SIGNals:SOURce:VOLTage<i>"
•
":POWer:SIGNals:SOURce:CURRent<i>"
NOTE
Use the lowest attenuation setting on the high voltage differential probes whenever possible
because the voltage levels on the deskew fixture are very small. Using a higher attenuation
setting may yield inaccurate skew values (and affect the measurements made) because the
noise level is magnified as well.
Содержание InfiniiVision M9241A
Страница 1: ...Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide ...
Страница 52: ...52 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 2 Setting Up ...
Страница 178: ...178 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 5 Commands Quick Reference ...
Страница 246: ...246 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 7 Root Commands ...
Страница 268: ...268 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 8 ACQuire Commands ...
Страница 280: ...280 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 9 CALibrate Commands ...
Страница 354: ...354 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 15 EXTernal Trigger Commands ...
Страница 412: ...412 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 17 FUNCtion m Commands ...
Страница 416: ...416 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 18 HCOPY Commands ...
Страница 442: ...442 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 20 MARKer Commands ...
Страница 588: ...588 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 23 MTESt Commands ...
Страница 692: ...692 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 24 POWer Commands ...
Страница 1062: ...1062 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 30 TIMebase Commands ...
Страница 1250: ...1250 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 33 WGEN w Commands ...
Страница 1306: ...1306 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 35 Obsolete and Discontinued Commands ...
Страница 1314: ...1314 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 36 Error Messages ...
Страница 1376: ...1376 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 39 More About Oscilloscope Commands ...