1364
Keysight InfiniiVision M9241/42/43A PXIe Oscilloscopes SCPI Programmer's Guide
38
Synchronizing Acquisitions
try:
KsInfiniiVisionX = rm.open_resource(VISA_ADDRESS)
except Exception:
print "Unable to connect to oscilloscope at " + str(VISA_ADDRESS)\
+ ". Aborting script.\n"
sys.exit()
# Set the Global Timeout
KsInfiniiVisionX.timeout = GLOBAL_TOUT
# Clear the instrument bus
KsInfiniiVisionX.clear()
# Clear all status registers and errors
KsInfiniiVisionX.write("*CLS")
try:
# Set up the oscilloscope
# ----------------------------------------------------------------
# Note that you would normally perform a reset (default setup) if
# you were to create the setup from scratch... But here we will
# use the oscilloscope "as is" for the most part.
# KsInfiniiVisionX.query("*RST;*OPC?")
# Resets the oscilloscope
# Always stop the oscilloscope when making any changes.
KsInfiniiVisionX.query(":STOP;*OPC?")
# For this example, the oscilloscope will be forced to trigger on
# the # (AC input power) LINE voltage so something happens.
# Always use normal trigger sweep, never auto.
KsInfiniiVisionX.write(":TRIGger:SWEep NORMal")
# This line simply gives the oscilloscope something to trigger on.
KsInfiniiVisionX.query(":TRIGger:EDGE:SOURce LINE;*OPC?")
# Clear the display (so you can see the waveform being acquired -
# otherwise, there is no need for this).
KsInfiniiVisionX.write(":CDISplay")
# Calculate acquisition timeout/wait time by short, overestimate
# method
# ----------------------------------------------------------------
# Create some default variables
N_AVERAGES = 1
N_SEGMENTS = 1
# Get some info about the scope time base setup
HO
= float(KsInfiniiVisionX.query(":TRIGger:HOLDoff?"))
T_RANGE
= float(KsInfiniiVisionX.query(":TIMebase:RANGe?"))
T_POSITION = float(KsInfiniiVisionX.query(":TIMebase:POSition?"))
# Determine Acquisition Type and Mode:
ACQ_TYPE = str(KsInfiniiVisionX.query(":ACQuire:TYPE?").strip("\n"))
ACQ_MODE = str(KsInfiniiVisionX.query(":ACQuire:MODE?").strip("\n"))
if ACQ_MODE == "SEGM":
Содержание InfiniiVision M9241A
Страница 1: ...Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide ...
Страница 52: ...52 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 2 Setting Up ...
Страница 178: ...178 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 5 Commands Quick Reference ...
Страница 246: ...246 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 7 Root Commands ...
Страница 268: ...268 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 8 ACQuire Commands ...
Страница 280: ...280 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 9 CALibrate Commands ...
Страница 354: ...354 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 15 EXTernal Trigger Commands ...
Страница 412: ...412 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 17 FUNCtion m Commands ...
Страница 416: ...416 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 18 HCOPY Commands ...
Страница 442: ...442 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 20 MARKer Commands ...
Страница 588: ...588 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 23 MTESt Commands ...
Страница 692: ...692 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 24 POWer Commands ...
Страница 1062: ...1062 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 30 TIMebase Commands ...
Страница 1250: ...1250 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 33 WGEN w Commands ...
Страница 1306: ...1306 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 35 Obsolete and Discontinued Commands ...
Страница 1314: ...1314 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 36 Error Messages ...
Страница 1376: ...1376 Keysight InfiniiVision M9241 42 43A PXIe Oscilloscopes SCPI Programmer s Guide 39 More About Oscilloscope Commands ...